It is important to consider the various information outlined to ensure appropriate pressure gauge selection and usage. Improper application can lead to instrument failure and possible personal injury and/or property damage. The following information is meant to be the overview of pressure gauge.
Types of Pressure Gauges
Winters offer a full range of pressure gauges suitable for
virtually all types of applications. Some specific types of
pressure gauges that will be covered in this manual are general-purpose
Bourdon tube, capsule-type, duplex, and differential pressure
gauges.
General-Purpose Bourdon Tube Gauge
T his common type of gauge utilizes a C-Bourdon tube, which
usually has an arc of 250 degrees. The process pressure is
connected to the fixed socket end of the tube, while the tip
end is sealed. Because of the difference between inside and
outside radii, the Bourdon tube tends to straighten when pressure
is applied. The non-linear motion at the tip of the Bourdon
tube is converted to linear rotational pointer response through
the use of gears and pinion movement. This type of pressure
gauge is used in medium to high-pressure measurements with
accuracies of 0.1 to 3% of full span. The pressure range capabilities
starts from 15 psi (100 kPa) to 20,000 psi (137.8 Mpa).
Capsule-Type Gauge
Capsule-type are formed from welding two pre-formed plates
together. Input process pressures cause the capsule to expand
or contract which is then translated into a linear rotational
pointer response through the use of gears and pinions. Bellows
gauges are typically used in low to intermediate and vacuum
pressure ranges with 0.1 to 1% of full span accuracy. Winters
Instruments capsule gauges have pressure ranges from 0-15" H2O (0-3.7 kPa) to 0-10 psi (0-68.9 kPa).
Duplex Gauge
A Duplex gauge indicates two related pressures in a process.
There are two Bourdon tubes inside the gauge casing which
results in two separate readings. The principal of the operation
is as described in the General-Purpose Bourdon Tube Gauge
section for Winters' Duplex Gauges.
Tire Pressure Gauge
There are 3 basic types of tire pressure gauges (also called tire gauges). The dial tire pressure
gauges, digital tire pressure gauges, and stick gauges. The gauges may look different, and display results differently, but all use the same principal. When the gauge is pressed onto the stem of a wheel, the air pressure in the tire pushes against some type of plunger which in turn actuates the readout section of the gauge assembly. Then we describe the similarities and differences between the different types of gauges.
Differential Gauges
The purpose of a differential gauge is similar to a duplex
gauge, but this gauge will indicate the difference between
two related pressures. Winters Instruments provide three different
types of differential gauges; parallel Bourdon tube, magnet
and piston, and opposing Bourdon tube designs.
Parallel Bourdon Tube Differential Gauge
The placement of the Bourdon tubes is similar to that of the
duplex gauges. The gears and pinions are calibrated to measure
the difference between the flex of the two Bourdon tubes and
this is translated to a single pointer. This type of differential
gauge is used when the indicating scale is 2/3 of the process
static pressure due to the limitations of the Bourdon tubes.
Piston Differential Gauge
The sensing element is a spring-biased diaphragm, which moves
linearly in proportion to the difference between two basic
pressures. A magnet on the HI pressure side of the piston
assembly moves with the diaphragm and rotates a follower magnet
located adjacent to the pressure cavity. The gauge pointer
is located at the end of the rotary magnet shaft and rotates
with the magnet to provide gauge readings proportional to
the differential pressures. This type of instrument can indicate
small values of differential pressure even when used in high-pressure
processes.
Return to Top
Pressure
Range
Select a pressure gauge with a full-scale pressure range of
approximately twice the normal operating pressure. For Bourdon
tube gauges, the maximum operating pressure should not exceed
75% of the full-scale range. The exception to this rule is
the opposing Bourdon tube differential gauge. Failure to select
a gauge within these criteria may result in premature fatigue
and failure of the Bourdon tube.
Temperatures
The normal ambient temperature ranges are -40C to 70C (-40F
to 158F) for dry gauges and -25C to 65C (-13F to 149F) for
glycerin filled gauges. The process temperature for dry gauges
can be up to 400C (752F). In situations where the process
temperature is extreme, utilizing a syphon or remote mounting
the pressure gauge with a capillary and diaphragm seal are
reasonable alternatives.In general, a gauge is unduly hot
if it cannot be grasped by hand without discomfort. It should
be noted that gauges used on water might burst if exposed
to frost. Please consult Winters for your specific requirements.
Materials
W inters provide a full range of casing and wetted parts to
accommodate all types of applications. The most common wetted
parts used are brass and 316 stainless steel. Brass wetted
parts are suitable for use on the following types of media;
air; oil, water and other non-corrosive fluids. For corrosive
fluids, stainless steel should be specified. For special applications,
other materials such as Monel can be used as well. Diaphragm
seals may also be considered. Please consult Winters Instruments
for assistance.
Vibrations
Vibrations and extreme ambient temperatures can affect the
dial reading of the pressure gauges. These areas should be
avoided as much as possible. Vibration effects can be minimized
by the use of a dampening liquid such as glycerin or silicone.
If vibration is extreme, then a flexible tube connected to
an appropriate diaphragm seal should be considered.
Pulsations
If pressures are expected to pulsate violently, oscillate
with high frequency or occur with sudden shock, a snubber
should be considered. Winters snubbers incorporate sintered
porous 316 stainless steel snubbing element with a large surface
area to ensure long term effectiveness on most pressure media.
Snubbers are available in the three viscosity classifications;
heavy oil, water and air. Brass, stainless steel or Monel
housing can be specified depending upon the media used.
Isolating Valves
It is recommended that isolating valves be fitted with the
gauge. This enables the gauge to be removed at any time for
checking, recalibration or replacing without interruption
to the process. The valves should be opened or closed slowly
to avoid sudden changes to pressure being applied to the gauge.
It is recommended for systems that have an abrupt pressure
surge at start-up to close the valves during initial start-up.
Installation Pressure Gauge
Location
V ibrations and extreme ambient temperatures can affect the
dial reading of the pressure gauge. These areas should be
avoided as much as possible. Vibration effects can be minimized
by the use of a dampening liquid such as glycerin or silicone.
If vibration is extreme, then a diaphragm seal connection
should be considered.
Mounting
A suitable thread sealant is required for N.P.T. threads such
as pipe
dope or Teflon tape. Never use any part of the pressure gauge
other than
the wrench flats that is on the gauge socket. Always tighten
with an open
end or adjustable wrench on the wrench flats. Failure to do
so will severely
damage the pressure gauge. Please note the following special
requirements
for the piston differential gauge.
Piston
Differential Gauge
Check instrument & identify HI and LO markings.
HI identifies the high-pressure port; LO the low pressure
port. If the
instrument is installed backwards, it will neither operate
nor be damaged.
It is recommended that the instrument be located above the
pressure source
to allow for drainage of the unit.
Please note that since there is a magnetic movement, the piston
differential gauge should never be mounted in direct
contact with a steel surface. Failure to do so will cause
a calibration
shift. Mount the gauge so that the gauge body is at least
1" (25.4 mm)
away from metal surface with non-magnetic spacers or an aluminum-mounting
bracket.
Venting
Procedures
Due to
pressure ¡°build up¡±, some gauges
(usually lower pressure ranges such as vacuum, up to 100 psi)
may reflect a reading that is slightly ¡°off zero¡±.
To properly ¡°vent¡± the pressure gauge,
please cut off tip of fill plug after you have installed the
instrument.
Return to Top
Operation
& Maintenance of Pressure Gauge
Storage
A ll pressure gauges should be placed in dry storage under
ambient room temperatures. It is recommended to store the
pressure gauges in their original packing boxes.
Inspection Frequency
W inters pressure gauges are rugged instruments featuring
simplicity of design to provide dependable and efficient service.
The frequency of inspection is dependent on the severity of
service and how critical the accuracy of the indicating pressure
is. The inspection frequency can range from monthly to annually
basis.
Return to Top
Recalibration
/ Assembly & Disassembly Procedures
Positive Pressure Gauge
Inspect the pressure gauge for "zero reading" when
there is no applied pressure. In many cases, the gauge pointer
can simply be adjusted and then the gauge will work within
specifications. However, if the gauge requires calibration,
then follow the outlines procedures below.
Remove the ring and lens
Slowly pressurize the gauge to its full scale and slowly release
the pressure back down to zero.
Check the gauge at a minimum of four equal pressure intervals
against a Certified Test Gauge.
Adjust the movement accordingly by removing the pointer with
a pointer extractor. Be careful not to bend the pointer shaft.
Removing the dial plate by loosening the holding screws.
When adjustments are complete, replace the dial plate and
pointer.
Re-check the gauge for accuracy.
Check lens assembly gasket for any deformation, tears or cracks.
If there are any visual defects, the gasket will need to be
replaced.
Re-assemble the lens assembly.
Capsule-Type Pressure Gauge
The same inspection procedure applies as per the positive
pressure gauge section above. The micrometer adjustable pointer
is accessible by removing the white plastic knob on the lens
and adjusting the screw. This is done without having to remove
the lens. If calibration is required, please follow the outlined
procedure below.
Remove lens by prying open the lens window with the notches
provided on the sides of the case.
Slowly pressurize the gauge to its full scale and slowly release
the pressure back down to zero.
Check the gauge at a minimum of four equal pressure intervals
against a Certified.
Test Gauge
Adjust the movement accordingly by removing the pointer with
a pointer extractor. Be careful not to bend the pointer shaft.
Remove the dial plate by loosening the holding screws.
When adjustments are complete, replace the dial plate and
pointer.
Re-check the gauge for accuracy.
Replace the lens window.
Piston Pressure Gauge.
Other than replacing broken lens, there is only one area where
the gauge may need attention. Erratic pointer action may indicate
that cleaning is required. Calibration is not required. However,
if the range spring is damaged or a new dial is required,
the gauge should be returned to the factory for parts and
recalibration.
For cleaning
Remove
the low pressure end cap with 1" (25.4 mm) spanner wrench.
Remove the range spring and the spacers at the bottom of the
spring pocket. BE CAREFUL NOT TO LOSE STACKING SPACERS, IF
ANY
Remove the four ?-20 socket head screws and separate the body
parts
Remove the diaphragm assembly
To replace the diaphragm, remove the screw at the top of the
diaphragm disc to separate the assembly. DO NOT REMOVE THE
SCREW IN THE MAGNET ASSEMBLY
Clean parts in a solvent solution after removing the "O"-ring
seals since solvents will attack the seal material
Place diaphragm disc on magnet shaft with "O"-ring
groove facing away from magnet. Place new diaphragm over shaft
with the open end of the convolution facing the magnet. Place
the second disc over diaphragm so that centre "O"-ring
groove faces diaphragm "O"-ring
Replace the spring cup, and the #8-32 assembly screw. Tighten
to finish assembly
Place the diaphragm assembly back into the diaphragm pocket
with the magnet end of the assembly at the high pressure side
Reassemble both body parts with the four ?-20 socket head
screw
Insert the spacers and the range spring into the spring pocket
of the end cap. While holding the instrument with the low
pressure port down, insert the end cap back into the low pressure
side of the body
Tighten the LO end cap and the instrument is now ready for
service
No recalibration is required.
For replacing the lens:
To replace a broken lens, check to see if the lens is held
on by a bezel or snap-ring. To remove a bezel, which is a
pressed on cover, either twist off by hand or pry off with
a screwdriver. To remove a snap-ring, pry out the ring with
a small screwdriver. Remove all glass chips, insert new lens
and re-insert the bezel or snap. With snap-rings, locate the
ring joint at the bottom of the gauge.
For replacing the pointer:
Remove bezel or snap-ring as previously described and clean
out glass chips. Remove old pointer with pointer extractor
or two small screwdrivers opposite each other under pointer
hub. Pry off evenly being careful not to bend the pointer
shaft. Install new pointer on zero. NOTE: Gauges with a zero
peg must have the pointer set at a reference pressure (preferable
mid-scale) to offset the preload against the zero pegs. Re-install
the lens, as described under lens replacement.
|